Acoustic measurement of the triple point of neon T (Ne) and thermodynamic calibration of a transfer standard for accurate cryogenic thermometry

Pan, CZ; Sparasci, F; Zhang, HY; Gambette, P; Plimmer, M; Imbraguglio, D; Gavioso, RM; Moldover, MR; Gao, B; Pitre, L

Pitre, L (corresponding author), LCM LNE Cnam, 61 Rue Landy, F-93210 La Plaine St Denis, France.; Gao, B; Pitre, L (corresponding author), Chinese Acad Sci, TIPCLNE Joint Lab Cryogen Metrol Sci & Technol, Beijing 100190, Peoples R China.; Pitre, L (corresponding author), Joint Res Lab Fluid Metrol Evangelista Torricelli, Paris, France.; Pitre, L (corresponding author), Joint Res Lab Fluid Metrol Evangelista Torricelli, Abano Terme, Italy.; Gao, B (corresponding author), Chinese Acad Sci, Tech In

METROLOGIA, 2021; 58 (4):

Abstract

We used absolute primary acoustic gas thermometry (AGT) to calibrate a Pt-Co resistance thermometer on the thermodynamic temperature scale by measurin......

Full Text Link